Catalog Number AE05ML
Device Problems Failure to Advance (2524); Device Fell (4014)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/19/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device history review for the product auto endo5 ml lot #73a1900236 investigation did not show issues related to the complaint.The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that during an operation, a clip got stuck in the applier and did not come out to be loaded properly.The user tried to load outside the patient's body several times, which resulted in the same occurrence.Therefore, the device was replaced with a new one.A clip fell but was retrieved; no clip remained in patient.
 
Manufacturer Narrative
(b)(4) after an additional medical review, a determination was made to re-classify the complaint as a malfunction.Complaint (b)(4) is being reported as a malfunction.There was no serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTO ENDO5 ML
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
MDR Report Key8995096
MDR Text Key157478652
Report Number3003898360-2019-01105
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
PMA/PMN Number
K152081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 08/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/06/2022
Device Catalogue NumberAE05ML
Device Lot Number73A1900236
Was Device Available for Evaluation? No
Date Manufacturer Received09/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-