C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER; FOLEY CATHETER (LATEX) Back to Search Results
Catalog Number 0165L16
Device Problems Deflation Problem (1149); Fluid/Blood Leak (1250); Device Fell (4014)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 08/21/2019
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the catheter fell out 2 days after the placement due to a water leakage.
 
Event Description
It was reported that the catheter fell out 2 days after the placement due to a water leakage.
 
Manufacturer Narrative
The reported issue was confirmed.The device was returned for evaluation.Visual inspection found a pinhole on the rubberized layer of the drainage lumen of the catheter, which caused an interlumen leak.The pinhole could possibly be created by latex bubble during the manufacturing process.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿ [warnings] 1.Method for use (1) do not inflate the balloon in the urethra.[the urethra may be injured.] (2) do not pull the catheter hard.[the bladder/urethra may be injured.] 2.Applicable patients (1) patients with delirium who might pull out catheter [when patient tugs at catheter unconsciously, the bladder and urethra may be damaged.]" correction: d4.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER
Type of Device
FOLEY CATHETER (LATEX)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8995779
MDR Text Key160209949
Report Number1018233-2019-05567
Device Sequence Number1
Product Code EZC
Combination Product (y/n)N
PMA/PMN Number
K910846
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 10/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2022
Device Catalogue Number0165L16
Device Lot Number9DA168
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2019
Initial Date Manufacturer Received 08/23/2019
Initial Date FDA Received09/11/2019
Supplement Dates Manufacturer Received09/25/2019
Supplement Dates FDA Received10/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-