Catalog Number 2400L
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 08/16/2019
Event Type  malfunction  
Manufacturer Narrative
The device was received for evaluation.A visual inspection was performed, and it was noted that the cable outer insulation has been torn open approximately seven of inches from the strain relief of the unit housing, exposing the ground braid; however, no actual wires or internal conductors exposed.Functional testing was performed with passing results with no errors or issues despite the damaged/frayed cable.The reported condition was verified.The cause of the reported condition was undetermined; however, is likely due to normal wear and tear.A device history review revealed no issues that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the load cell cable of an automated compounding device (acd) was damaged, exposing wires which were seen during setup/preparation.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EM 2400, LOAD CELL MODULE
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE - ROUND LAKE
round lake IL
Manufacturer (Section G)
BAXTER HEALTHCARE - ROUND LAKE
25212 w. illinois route 120
round lake IL 60073
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9009923
MDR Text Key159044481
Report Number1416980-2019-04991
Device Sequence Number1
Product Code NEP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 09/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/12/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2400L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/16/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-