It was reported that the tip of the wire was torn.A 190 cm filterwire ez was selected for use in the internal carotid and common carotid.During procedure, the physician noted discomfort on the tip of the wire when the device was attempted to approach in to the distal of the stenosis area, so it was removed.The resistance was noticed when the device was inserted into the patient's body from the guiding catheter.Subsequently, it was noted that the tip of the wire was torn.The device was replaced with another of the same device to successfully complete the procedure.No patient complications were reported and the patient condition was good.
|