• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2272
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Twiddlers Syndrome (2114)
Event Date 08/22/2019
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient had a pacemaker check on (b)(6) 2019 after the initial implant procedure.During the device interrogation, it was discovered that the right ventricular lead exhibited intermittent undersensing and intermittent capture.It was noted that the patient had been twiddling with the pacemaker in the implant pocket.A chest x-ray was performed which showed that the pacemaker had migrated down and the slack of both the atrial and right ventricular leads had decreased considerably.The device and leads were revised successfully.The patient was asymptomatic before, during, and after the revision procedure.Related manufacturer reference number: 2017865-2019-13432, 2017865-2019-13684.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY MRI
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key9013240
MDR Text Key158163773
Report Number2017865-2019-13685
Device Sequence Number1
Product Code DXY
UDI-Device Identifier05414734509589
UDI-Public05414734509589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/12/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2021
Device Model NumberPM2272
Device Lot NumberA000082814
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/22/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/17/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age88 YR
Patient Weight102
-
-