Model Number 27MECJ-502 |
Device Problem
Obstruction of Flow (2423)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/15/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
|
Event Description
|
On (b)(6) 2017, a 27mm masters series valve was implanted.On (b)(6) 2019, the valve was explanted due to leaflet obstruction and the valve was exchanged with a 27mm epic stented porcine valve.The patient has been discharged and no patient consequences were reported.
|
|
Manufacturer Narrative
|
The reported event of a "stuck" valve could not be confirmed.The investigation found that both leaflets opened and closed completely and easily.Pannus was present on the remains of the sewing cuff but did not impinge or obstruct the leaflets.No acute inflammation or significant calcifications were present.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.The cause of the reported event could not be conclusively determined.
|
|
Search Alerts/Recalls
|