The initial reporter complained of questionable thyroid results for 1 patient sample on a cobas 8000 e 801 module.The patient sample was submitted for investigation where discrepant results were identified for elecsys tsh assay, elecsys ft4 iii assay, and elecsys ft3 iii between the customer's e 801 module and an e 801 module used at the investigation site compared to the siemens centaur method used at the investigation site.It was asked but not known if the results from the customer site were reported outside of the laboratory.This medwatch will cover ft4 iii.Refer to medwatch with patient identifier (b)(6) for information on the tsh results and medwatch with patient identifier (b)(6) for information on the ft3 iii results.The customer's cobas e 801 serial number (b)(4).
|
All values, generated with the analyzers from roche diagnostics and the analyzer from siemens, are within the normal reference ranges of the respective assays.The observed differences in ft4 values generated with the roche assay and siemens centaur assay are most likely caused by differences in the overall setups of the assays, the antibodies used, differences in reference materials, and the differences in the standard.
|