• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC BATTERY HANDPIECE/MODULAR FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC BATTERY HANDPIECE/MODULAR FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 05.001.201
Device Problem Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/04/2019
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).Concomitant medical devices: lid device, chuck device, ao reaming attachment device, quick coupling device, (b)(6) 2019.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and / or duplicated.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
This is report 1of 4 for the same event: it was reported from (b)(6) that the battery handpiece / modular device, chuck device, ao reaming attachment device and quick coupling device had heating issues while being used together with the lid device.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Correction: manufacturing site name: the manufacturer location was incorrectly documented as (b)(6) in the initial report.The location has been updated to (b)(6).The contact office name/address has been updated accordingly to reflect the corrected manufacturing facility.Device history record review: a device history review was performed and no non-conformances were detected related to the reported condition.Device evaluation: the actual device was returned for evaluation.During repair, it was determined that the reported condition was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was determined that the bearings were worn.It was further determined that the device failed pretest for check of free moving, check proper function of the triggers and check history.The assignable root cause was determined to be traced to component failure (faulty parts) , which is normal wear.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY HANDPIECE/MODULAR FOR TRS
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key9086195
MDR Text Key164312699
Report Number8030965-2019-68403
Device Sequence Number1
Product Code GEY
UDI-Device Identifier07611819977815
UDI-Public07611819977815
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2019
Date Manufacturer Received11/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-