Model Number N/A
Device Problem Mechanical Problem (1384)
Patient Problem No Code Available (3191)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Concomitant medical products: catalog number:010000846 lot number: 6551473 brand name:g7 neutral liner.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2019-04143.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during an initial hip arthroplasty , the liner would not assemble with the shell resulting in an intraoperative delay of approximately 30 to 60 minutes.The surgery was completed with a back up device.Additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.  reported event was unable to be confirmed due to limited information received from the customer.Device history record was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Device not returned for evaluation.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 OSSEOTI MULTIHOLE 46MM B
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9088056
MDR Text Key160518861
Report Number0001825034-2019-04145
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
PMA/PMN Number
K140669
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110010261
Device Lot Number6401304
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-