SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID, PRODUCT CODE: LGZ
|
Back to Search Results |
|
Catalog Number CON-HL-90 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Event Description
|
Information was received that a smiths medical level 1 hotline blood and fluid warmer had a "leaking from front cover".No adverse effects were reported.
|
|
Manufacturer Narrative
|
One fluid warmer was returned for evaluation.Visual inspection of the device found it to be in fair condition.Device underwent functional testing by filling tank with water and performed power up of the device.As a result, it was noted to be leaking through the enclosure, confirming the reported customer complaint.The problem source was determined to be disconnection of the hose causing the water to leak.This investigation revealed no intrinsic evidence to suggest a cause of issue related to manufacturing.
|
|