• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 5.0FR URETHANE UMB CATH; CATHETER, UMBILICAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 5.0FR URETHANE UMB CATH; CATHETER, UMBILICAL ARTERY Back to Search Results
Model Number 8888160341
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/13/2019
Event Type  malfunction  
Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation. if the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted. as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
 
Event Description
The customer reported that the argyle umbilical catheter broke into two pieces when the rn was removing it.There was no patient injury.
 
Manufacturer Narrative
A device history record review could not be performed because a lot number was not received with the complaint.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to the release of product.One decontaminated sample was received at the manufacturing site for evaluation.Upon a visual evaluation of the sample, the reported condition was confirmed.A definitive root cause could not be determined at this time.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling) are in place to prevent nonconforming product from leaving the manufacturing operations.A corrective action is not applicable at this time.Functional testing and visual inspections are being performed according to our current quality standards and inspection procedures.This complaint will be used for qa tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0FR URETHANE UMB CATH
Type of Device
CATHETER, UMBILICAL ARTERY
Manufacturer (Section D)
COVIDIEN
edificio b20 calle #2 zona fra
alajuela
CS 
MDR Report Key9100767
MDR Text Key163307345
Report Number3009211636-2019-00672
Device Sequence Number1
Product Code FOS
UDI-Device Identifier20884527005120
UDI-Public20884527005120
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number8888160341
Device Catalogue Number8888160341
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/17/2019
Initial Date FDA Received09/20/2019
Supplement Dates Manufacturer Received09/17/2019
Supplement Dates FDA Received01/24/2020
Patient Sequence Number1
Patient Age7 DA
-
-