Reportable based on device analysis completed on (b)(6) 2019.It was reported that the device could not cross the lesion.The 80% stenosed, 32x3.0mm target lesion was located in a severely tortuous and moderately calcified left circumflex artery (lcx).A 10/2.50 flextome cutting balloon was selected for use.During the procedure, it was noted that the device was unable to cross the lesion.The procedure was completed with another of the same device.No complications reported and the patient's condition was stable.However, device analysis revealed that the device was leaking due to a balloon pinhole located at the proximal end of the balloon.
|