Model Number G148 |
Device Problems
Premature Discharge of Battery (1057); Device Displays Incorrect Message (2591)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/05/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
This product has been received for analysis.This report will be updated upon completion of analysis.
|
|
Event Description
|
It was reported that this device exhibited premature battery depletion associated with a recorded code 1003 indicative of battery voltage too low for the projected remaining capacity.Data analysis showed the battery appeared to be depleting more quickly than expected.This device was explanted and returned for analysis.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the product was performed.Review of device memory confirmed the presence of a higher than normal current drain condition.Electrical testing and analysis were then conducted, which isolated the high current to an anomaly in an oxide layer within a custom integrated circuit component.This anomaly caused a high current drain, which over time resulted in the reported clinical observations.
|
|
Event Description
|
This supplemental report is being filed to include product investigation details.
|
|