• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN X4 CRT-D; IMPLANTABLE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN X4 CRT-D; IMPLANTABLE DEVICE Back to Search Results
Model Number G148
Device Problems Premature Discharge of Battery (1057); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/05/2019
Event Type  Injury  
Manufacturer Narrative
This product has been received for analysis.This report will be updated upon completion of analysis.
 
Event Description
It was reported that this device exhibited premature battery depletion associated with a recorded code 1003 indicative of battery voltage too low for the projected remaining capacity.Data analysis showed the battery appeared to be depleting more quickly than expected.This device was explanted and returned for analysis.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the product was performed.Review of device memory confirmed the presence of a higher than normal current drain condition.Electrical testing and analysis were then conducted, which isolated the high current to an anomaly in an oxide layer within a custom integrated circuit component.This anomaly caused a high current drain, which over time resulted in the reported clinical observations.
 
Event Description
This supplemental report is being filed to include product investigation details.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN X4 CRT-D
Type of Device
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key9105574
MDR Text Key159806609
Report Number2124215-2019-19010
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public00802526534591
Combination Product (y/n)N
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 01/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/20/2019
Device Model NumberG148
Device Catalogue NumberG148
Device Lot Number163803
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2019
Initial Date Manufacturer Received 09/09/2019
Initial Date FDA Received09/23/2019
Supplement Dates Manufacturer Received10/07/2019
Supplement Dates FDA Received01/28/2020
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age84 YR
-
-