The lens was returned for evaluation.Solution is dried on the lens.The lens is pressed against three posts in the lens case.One haptic tip is pointing down into a drain hole in the lens case.The lens was cleaned with lphse.A light scratch is observed on the anterior side of the optic surface.The scratch may be interpreted as the reported damage of a crease in the middle of the optic.Product history records were reviewed and the documentation indicated the product met release criteria.Associated products were not provided.It is unknown if qualified products were used.The reported optic damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The manufacturer internal reference number is: (b)(4).
|