• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SA60AT
Device Problem Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/06/2019
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.Additional information has been requested.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported there was a crease in the middle of the optic on an intraocular lens (iol).There was no reported patient impact and the procedure was completed using a backup lens.Additional information was requested.
 
Manufacturer Narrative
The lens was returned for evaluation.Solution is dried on the lens.The lens is pressed against three posts in the lens case.One haptic tip is pointing down into a drain hole in the lens case.The lens was cleaned with lphse.A light scratch is observed on the anterior side of the optic surface.The scratch may be interpreted as the reported damage of a crease in the middle of the optic.Product history records were reviewed and the documentation indicated the product met release criteria.Associated products were not provided.It is unknown if qualified products were used.The reported optic damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF SINGLEPIECE IOL
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
cindy milam
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152231
MDR Report Key9106248
MDR Text Key165641767
Report Number1119421-2019-01565
Device Sequence Number1
Product Code HQL
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Model NumberSA60AT
Device Catalogue NumberSA60AT.190
Device Lot Number12602380
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2019
Date Manufacturer Received11/17/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-