Model Number 7880 |
Device Problems
Detachment of Device or Device Component (2907); Material Integrity Problem (2978)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/03/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that catheter separation occurred.The target lesion was located in the aorta.A 5f 145 110 cm impulse guide catheter was selected for use during procedure.It was observed that the impulse catheter separated at the 145 degree bend.Subsequently, a snare was advanced through the femoral access and the detached catheter's tip was retrieved.The device was completely removed from the patient's body.There were no patient complications reported.Patient is stable.
|
|
Event Description
|
It was reported that catheter separation occurred.The target lesion was located in the aorta.A 5f 145 110cm impulse guide catheter was selected for use during procedure.It was observed that the impulse catheter separated at the 145 degree bend.Subsequently, a snare was advanced through the femoral access and the detached catheter's tip was retrieved.The device was completely removed from the patient's body.There were no patient complications reported.Patient is stable.
|
|
Manufacturer Narrative
|
(b)(6).Device evaluated by mfr: the device was returned in two pieces.Analysis of the tip and shaft included microscopic, visual and sem inspection.Inspection revealed a complete separation of the shaft located 5.5 cm from the distal end.The separated ends were run through an electron microscope and the edge of the distal separated edge found full bond edge with evidence of tensile forcep applied to device.Inspection of the rest of the device found no other damage or defect.
|
|
Search Alerts/Recalls
|