• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT QUARTET LEAD SMALL-S, 86 CM; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT QUARTET LEAD SMALL-S, 86 CM; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number 1456Q/86
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Vascular Dissection (3160)
Event Date 09/11/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
During the initial implant procedure, a dissection of the coronary sinus was noted from the left ventricular (lv) lead.The procedure was aborted and a new lv lead was implanted successfully at a later date.The patient was stable.
 
Manufacturer Narrative
As received, a complete lead was returned in one piece.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual inspection or x-ray examination of the lead did not find any anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD SMALL-S, 86 CM
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key9112132
MDR Text Key159870528
Report Number2017865-2019-14133
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510158
UDI-Public05414734510158
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2021
Device Model Number1456Q/86
Device Catalogue Number1456Q/86
Device Lot NumberA000058168
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/11/2019
Initial Date FDA Received09/24/2019
Supplement Dates Manufacturer Received11/12/2019
Supplement Dates FDA Received11/12/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/21/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age65 YR
-
-