Model Number 42856
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the valve had poor drainage.The valve was replaced, and the new valve was working normally.
 
Manufacturer Narrative
The pl setting on the returned valve was observed to be between 2.5 and 0.5.The laboratory personnel could not adjust the pl setting on the valve prior to or after flushing.Therefore, the valve could not be tested for valve flux, patency, leakage, siphon, reflux, pressure/flow and pre-implantation tests.It was observed that water was able to flow through the valve during flushing.Biological debris were observed on the interior of the valve.The instructions for use cautions, ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization or other debris.¿ all valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATA II VALVE, SMALL
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer (Section G)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer Contact
katcha taylor
9775 toledo way
irvine, CA 92618
7635263305
MDR Report Key9120284
MDR Text Key161737783
Report Number2021898-2019-00346
Device Sequence Number1
Product Code JXG
UDI-Device Identifier00643169975828
UDI-Public00643169975828
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K042465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2020
Device Model Number42856
Device Catalogue Number42856
Device Lot NumberE41719
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/24/2019
Initial Date FDA Received09/25/2019
Supplement Dates Manufacturer Received12/19/2019
Supplement Dates FDA Received12/26/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Weight51
-
-