• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JABIL CIRCUIT(SHANGHAI) LTD INVOS; OXIMETER, TISSUE SATURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JABIL CIRCUIT(SHANGHAI) LTD INVOS; OXIMETER, TISSUE SATURATION Back to Search Results
Model Number DS5100C-PB
Device Problem Low Readings (2460)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the device had intermittent low baselines on healthy patients.There was no patient injury.
 
Manufacturer Narrative
H3 evaluation summary: one sample was received for evaluation.Service personnel reported that could not be able to duplicate the issue.The device passed the all test criteria.The investigation found the device to function normally.Information has been added to the database and trends will continue to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVOS
Type of Device
OXIMETER, TISSUE SATURATION
Manufacturer (Section D)
JABIL CIRCUIT(SHANGHAI) LTD
no 600 tian lin rd
shanghai 20023
CN  20023
MDR Report Key9121207
MDR Text Key160209502
Report Number2936999-2019-00787
Device Sequence Number1
Product Code MUD
UDI-Device Identifier10884521185906
UDI-Public10884521185906
Combination Product (y/n)N
PMA/PMN Number
K091224
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 11/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDS5100C-PB
Device Catalogue NumberDS5100C-PB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2019
Initial Date Manufacturer Received 08/30/2019
Initial Date FDA Received09/25/2019
Supplement Dates Manufacturer Received10/24/2019
Supplement Dates FDA Received11/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-