Evaluation summary: one device was received for evaluation.The returned sample met specification as received.The visual inspection found there were no indications of transport damage, one gaiter was chipped and the wheels were scratched.The reported condition was not confirmed.The investigation found the device to function normally and within specifications.All replacement parts were installed, the device was fully tested and is functioning per specifications.Calibration and testing were conducted.If information is provided in the future, a supplemental report will be issued.
|