Device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (connector damaged) has been confirmed.Upon investigation the trunk cable connector was physically damaged and the electrode belt was unable to securely connect to a monitor.Additionally, the belt failed a functionality test.The cause for the test failure was missing heat shrink tubing on the pulse wires inside the distribution node (dn).The root cause for the damaged connector was excessive force.The root cause for the missing heat shrink was unable to be positively identified.No adverse event resulted from the defective electrode belt.
|