ST. JUDE MEDICAL - NEUROMODULATION SWIFT-LOCK ANCHOR; SCS ANCHOR Back to Search Results
Model Number 1192
Device Problems Fracture (1260); High impedance (1291)
Patient Problem Inadequate Pain Relief (2388)
Event Date 09/09/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
Related manufacturer reference number: 1627487-2019-10924, related manufacturer reference number: 1627487-2019-10925, related manufacturer reference number: 1627487-2019-10926.It was reported that the patient was experiencing autoreducing.An impedance check showed that all 16 contacts were over 3000 ohms.A fluoro lead check showed no lead fractures or disconnections.The patient denied any falls or trauma.As a result, a revision was performed where leads and anchors were explanted and replaced.During the procedure, it was discovered that a swiftlock anchor was torn into 2 pieces and the lead had fractured at the site where swiftlock was anchored to it.Therapy was restored following the procedure.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWIFT-LOCK ANCHOR
Type of Device
SCS ANCHOR
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key9129067
MDR Text Key160347889
Report Number1627487-2019-10927
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067024084
UDI-Public05415067024084
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2020
Device Model Number1192
Device Lot Number6376146
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/09/2019
Initial Date FDA Received09/27/2019
Supplement Dates Manufacturer Received09/27/2019
Supplement Dates FDA Received10/02/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/29/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS ANCHOR; SCS LEAD X 2
Patient Outcome(s) Other;
Patient Weight88
-
-