The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported material separation.Both, the delivery system tip and the supported mandrel tip are white, and they are next to each other at the distal end of the device.There is no indication of a product quality issue with respect to manufacture, design or labeling.Exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.
|
It was reported that during unpacking of a rx.014 acculink 6-8/40, after taking out the stent from the dispenser hoop coil, as the stylet was being removed, the white cone [catheter tip] on top of the stent came off with the stylet.The device was not used and there was no patient involvement.A new device was replaced to successfully complete the procedure.There was no clinically significant delay or adverse patient effects.No additional information was provided.
|