• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH CUP IMPACTOR BOLT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH CUP IMPACTOR BOLT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 1440-2011
Device Problem Separation Problem (4043)
Patient Problem Bone Fracture(s) (1870)
Event Date 07/25/2019
Event Type  Injury  
Manufacturer Narrative
If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Not returned.
 
Event Description
When impacting a trident psl 50 cup, the magnetised offset cup impactor bolt (1440-2011) got stuck in the definitive cup & couldn¿t be detached.The surgeon had to dislodge the cup, take it out, ream up and implant a 52 psl cup with a different impactor (the standard trident cup impactor).Upon dislodging the cup to take it out, the anterior column fractured.Approx 5-10 minutes surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUP IMPACTOR BOLT
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
Manufacturer Contact
brad curtis
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key9131169
MDR Text Key165582879
Report Number0002249697-2019-03341
Device Sequence Number1
Product Code LPH
UDI-Device Identifier07613327196245
UDI-Public07613327196245
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K161569
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial
Report Date 09/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number1440-2011
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/25/2019
Initial Date FDA Received09/27/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age82 YR
Patient Weight69
-
-