• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 STR MONOBLOCK SHELL INSRTR; INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 STR MONOBLOCK SHELL INSRTR; INSTRUMENT Back to Search Results
Model Number N/A
Device Problem Material Disintegration (1177)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/01/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Concomitant medical products: 110003450 g7 str monoblock shell insrtr 280280.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2019-04173.
 
Event Description
It was reported that during surgery, the threads on the inserter were damaged.Went to use the back-up unit and those threads were damaged also.Surgery was completed without delay or consequence by using an offset inserter.Attempts were made to obtain additional information; however, none was available.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Updated: b4, b5, g4, h2, h3, h6.Complaint sample was evaluated and the reported event was confirmed.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.Both g7 inserters were returned and evaluated against the complaint.Visual inspection confirmed one of the threaded tips to be fractured such that a portion of the face of the tip remains on the inserter, while the other one fractured completely.Dings and scratches were observed on both haft.Impact marks were found on the both strike plate consistent with a multiple use device.Fracture analysis confirmed the fractures were due to bending overload of the threaded tip.The material was confirmed to be 455 ss.Devices had an estimated field age of 5 and 3 years, respectively.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR MONOBLOCK SHELL INSRTR
Type of Device
INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9139316
MDR Text Key160902414
Report Number0001825034-2019-04312
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 02/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110003450
Device Lot Number791450
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/17/2019
Initial Date FDA Received10/01/2019
Supplement Dates Manufacturer Received01/31/2020
Supplement Dates FDA Received02/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-