Model Number 6660 |
Device Problem
Use of Device Problem (1670)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/19/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results, method, and conclusion codes along with the investigation results will be provided in final report.
|
|
Event Description
|
Related manufacturer reference number: 1627487-2019-11074, related manufacturer reference number: 1627487-2019-11076.It was reported during an ipg replacement procedure for normal end of life the physician was unable to remove the extensions from the ipg and damaged the ipg header.The physician attempted to insert the extensions into a new ipg.The extensions could not be fully inserted so the extensions were explanted and replaced resolving the issue.
|
|
Manufacturer Narrative
|
The reported event the ipg was depleted was confirmed.As received, the ipg displayed the ¿replace generator ¿.The uep inductive tool showed an eri time stamp had been logged.The device was running the therapy application.Once the eri was set to zero, ble communication could be established.A longevity calculation confirmed normal battery depletion based on average device usage using the patient parameters provided.The reported extension insertion and removal issue was not confirmed.The ipg was returned with extensive damage to port 9-16 of the header.The electrodes were unseated and multiple feed through wires were broken.The damage observed is consistent with the device being subjected to an overstress condition during the explant procedure.This is consistent with the event details that reported the ¿header got broken with forceps¿.As the condition of the ipg prior to explant could not be ascertained, the allegation the extensions could not be inserted or removed from the header could not be conclusively determined.
|
|
Event Description
|
Related manufacturer reference number: 1627487-2019-11074.Related manufacturer reference number: 1627487-2019-11076.
|
|
Search Alerts/Recalls
|