The cpc (colder products company) connector is a component that provides the interface between the driver drivelines and the tah-t cannula.The customer, a syncardia certified hospital, reported that the spring in the freedom drivers' cpc connector was broken.There was no reported adverse patient impact.The customer also reported that the patient was subsequently switched to a backup freedom driver.
|
During the course of the investigation, it was discovered that the tah-t cpc connector was attached to the freedom driver driveline and that was the cpc connector with the issue.Therefore, the appropriate fields in sections b and d have been corrected to reflect the change in product from freedom driver to tah-t.Visual inspection could not confirm the customer-reported issue of a broken spring, as the spring was missing from within the housing, beneath the metal thumb release tab.It is likely that a broken or displaced spring could result in the spring becoming completely dislodged and missing from beneath the metal thumb release tab.Clinicians and patients are trained to thread a wire tie beneath the thumb release tab of the cpc connectors to prevent accidental disconnection of the cannulae from the drivelines.Although a definitive root cause for the displaced spring could not be determined, it is possible that a cpc connector spring can be displaced when the wire tie is threaded through, or removed from, the connector during or after a driver switch or during driver training.Syncardia has an open corrective and preventive action (capa) to address this issue with cpc connectors.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
|