MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Patient Involvement (2645)
Event Date 09/12/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses low risk to a patient because the issue was observed when the freedom driver was not supporting a patient.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The freedom driver was not supporting a patient.The customer, a syncardia authorized distributor, reported that the patient's backup freedom driver display showed asterisks instead of numbers for the fill volume and cardiac output.
 
Manufacturer Narrative
During investigational testing, the customer-reported issue of the driver display showing asterisks instead of numbers for the cardiac output and fill volume was confirmed and reproduced.The root cause was determined to be a malfunction of the airflow sensor which was found to have internal wire corrosion.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 4965 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key9141514
MDR Text Key161040236
Report Number3003761017-2019-00283
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 11/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2019
Initial Date Manufacturer Received 09/12/2019
Initial Date FDA Received10/01/2019
Supplement Dates Manufacturer Received09/12/2019
Supplement Dates FDA Received11/07/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age58 YR
-
-