|
Catalog Number 208370401 |
Device Problems
No Audible Alarm (1019); Improper Flow or Infusion (2954)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/09/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is expected to return to the manufacturer for investigation; it has not been received.
|
|
Event Description
|
The event involved a report stating that a large amount of air was seen in the iv line of a pca pump and the pump did not alarm.The medication involved in the event was dilaudid.The customer further stated that the documentation on the pump stated that there was 6.3mg of diluadid, when the max amount is 6mg.There was a patient involved, however, there was no serious injury, no adverse event, and no medical intervention required.
|
|
Manufacturer Narrative
|
H10: the device was received by the manufacturer on (b)(6) 2019.The device history was reviewed and found no customer protocol; history was cleared on 10/17/19.There was no protocol recorded in the history log on the event date 09/09/19.During testing, the device was programmed at the following setting: drug concentration: 1 mg/ml, mode: pca only, pca dose: 2 mg, pca lockout: 5 minutes, 4 hours dose limit: 6 mg.After infusing for 3 pca doses, the reading was 6 mg on the pump display.Actual volume measured on the scale (d5547) is 6.0 ml.The error rate was 0.00% (tsm-limit +/- 5%).Device delivered dose properly when patient pendant button is pushed.There is no self-delivery observed during testing.The device also passed pvt delivery accuracy test at 20.15 ml (spec 19 to 21 ml).This device did not have air sensor to detect air in line, therefore air in line will not alarm when air in line is present.The customer complaint for ¿air was in the iv line of the pca and no alarm sounded and the documentation on the pump stated that there was 6.3 mg of bialubib when the max amount is only 6 mg.There was air in the line but the pca is not doing an alarm¿ could not duplicate during testing.Probable cause: no probable cause was found as the device passed delivery accuracy test within the specification.
|
|
Search Alerts/Recalls
|
|
|