Model Number R SERIES |
Device Problem
Pacing Problem (1439)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during biomed testing, the device displayed "pacer fault 117" and "pacer warning 124" messages.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical (b)(4) for evaluation.The reported malfunction was duplicated, and the pace/defib engine board was replaced to remedy the problem.The device was rectified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
|
Manufacturer Narrative
|
The device was not returned to zoll medical corporation; the device's pace/defib engine board was removed and returned.The customer's report was duplicated and attributed to a faulty resistor on the pace/defib engine board.Analysis for reports of this type has not identified an increase in trend.
|
|