• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DYNAGEN CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DYNAGEN CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G150
Device Problems Use of Device Problem (1670); Device Displays Incorrect Message (2591); Data Problem (3196)
Patient Problem No Code Available (3191)
Event Date 08/27/2019
Event Type  Injury  
Manufacturer Narrative
This investigation will be updated should further information be provided.
 
Event Description
It was reported that this device had been programmed off during a surgical procedure and was programmed back on after the procedure.However, a red alert had been generated with information that the device was programmed off.A boston scientific technical services consultant reviewed the stored data and revealed that the device had been programmed off for the surgery, programmed on after the surgery and then programmed off again.Device therapy was programmed back on.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key9147926
MDR Text Key161392434
Report Number2124215-2019-19793
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534614
UDI-Public00802526534614
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/01/2021
Device Model NumberG150
Device Catalogue NumberG150
Device Lot Number494309
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/28/2019
Initial Date FDA Received10/02/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/09/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age69 YR
-
-