There has been a previous report received where this malfunction has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.Please note that while this product is not sold in the us, it is considered similar to products that are marketed in the us by dentsply sirona.The device does what it is supposed to do, but the factory settings were switched and permanently adjusted (saved).For the file system protaper next, autorev was from auto reversing adjusted to auto stop.Function: auto reversing: automatic stop and left turn when the preset torque is reached, followed by clockwise rotation.Auto stop: automatic stop and left turn when the preset torque is reached, followed by stop.A permanent change of the direction of rotation e.G.Counterclockwise rotation is only possible via the rev key.Device has been reset to factory settings and calibrated.
|