• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FILTERWIRE EZ; TEMPORARY CAROTID CATHETER FOR EMBOLIC CAPTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FILTERWIRE EZ; TEMPORARY CAROTID CATHETER FOR EMBOLIC CAPTURE Back to Search Results
Model Number 44150
Device Problems Difficult or Delayed Positioning (1157); Fracture (1260); Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/24/2019
Event Type  malfunction  
Event Description
It was reported that wire fracture occurred.The 80% stenosed target lesion was located in the severly tortuous internal carotid artery.A 190cm filterwire ez was selected for use in calcified aortic stenosis (cas) procedure.During procedure, resistance was felt when the filter wire was deployed several times.Upon removal of the wire, it was noted that wire part was broken.The device was discontinued and removed.The procedure was completed with another of same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FILTERWIRE EZ
Type of Device
TEMPORARY CAROTID CATHETER FOR EMBOLIC CAPTURE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9152526
MDR Text Key161530167
Report Number2134265-2019-11921
Device Sequence Number1
Product Code NTE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K063313
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/05/2020
Device Model Number44150
Device Catalogue Number44150
Device Lot Number0022899413
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/16/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/06/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age80 YR
-
-