BECTON DICKINSON HUNGARY KFT (BD) ULTRASAFE X100L PNG CLEAR NVS STEIN; PISTON SYRINGE Back to Search Results
Catalog Number 47439230
Device Problem Failure to Deliver (2338)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/04/2019
Event Type  malfunction  
Manufacturer Narrative
Pma/50(k)#: k011369, k122558.Unconfirmed, no issue observed.The customer issued a complaint for pre-activated device detected by end user.Neither sample nor photo was provided to bd medical ¿ pharmaceutical system (bdm-ps) for analysis.Bdm-ps performed a batch history record¿s review (bhr) including a review of all data collected during in process and quality inspections.The batches involved in this complaint meet all acceptable quality levels (aql¿s), were manufactured and released according to applicable procedures and specifications.Based on investigation conclusion, bdm-ps was not able to confirm the symptom perceived by customer or correlate this symptom with a potential cause linked to bd process.During the years of investigation of complaints about pre-activated devices several root causes were discovered which were within the customer¿s sphere of influence.Best practices and guidelines were collected and summarized.Rationale: complaint is unconfirmed.
 
Event Description
It was reported that 2 ultrasafe x100l png clear nvs stein experienced a broken syringe syringe/pen which was noted prior to use.The following information was provided by the initial reporter: consumer called to report that the needles in her syringes popped into the safety shields as soon as they were taken out of the box and that one of them even completely unattached itself from the syringe.
 
Manufacturer Narrative
The following information has been corrected: d.3.Reason code for no evaluation: other.D.3.If other specify: see h.10 h3 other text : see h.10.
 
Event Description
It was reported that 2 ultrasafe x100l png clear nvs stein experienced a broken syringe syringe/pen which was noted prior to use.The following information was provided by the initial reporter: consumer called to report that the needles in her syringes popped into the safety shields as soon as they were taken out of the box and that one of them even completely unattached itself from the syringe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASAFE X100L PNG CLEAR NVS STEIN
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON HUNGARY KFT (BD)
uveggyar utca 3
kornye tatabanya 2851
HU  2851
MDR Report Key9153237
MDR Text Key195108833
Report Number3009081593-2019-00225
Device Sequence Number1
Product Code MEG
Combination Product (y/n)N
PMA/PMN Number
K011369
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 10/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/30/2022
Device Catalogue Number47439230
Device Lot Number7117213
Date Manufacturer Received09/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-