ABBOTT DIABETES CARE INC FREESTYLE LIBRE 14 DAY; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71940-01
Device Problem Low Readings (2460)
Patient Problems Headache (1880); Hyperglycemia (1905); Vertigo (2134)
Event Date 09/07/2019
Event Type  Injury  
Manufacturer Narrative
No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The reported complaint does not pertain to the freestyle libre reader.Therefore no further investigation into the reader is required.Dhrs for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.All review activities conducted above, including but not limited to the final release testing specifically associated with the manufacture of this product, are sufficient information in order to show if the product has met specifications prior to release.If the product is returned, the case will be re-opened and a physical investigation will be performed.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A low readings issue was reported with the adc freestyle libre sensor.Customer reported receiving a message 'lo' (reading less than 40 mg/dl) on the sensor and experiencing headache and vertigo.Customer self-treated with novorapid insulin (dose unknown) and called emergency services, who transported him to a hospital.At the hospital, a reading of 800 mg/dl was obtained on the hospital device and customer was diagnosed with hyperglycemia and treated with "3 perfusions to rehydrate" and insulin.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 14 DAY
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
devin dirstine
1360 south loop road
alameda, CA 94502-7001
5108644391
MDR Report Key9153926
MDR Text Key161177930
Report Number2954323-2019-07623
Device Sequence Number1
Product Code PZE
Combination Product (y/n)N
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial
Report Date 10/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71940-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/09/2019
Initial Date FDA Received10/03/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/09/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age42 YR
Patient Weight95
-
-