Model Number MAJ-1500
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/10/2019
Event Type  malfunction  
Manufacturer Narrative
The subject maj-1500 has not been returned to olympus medical systems corp.(omsc) for evaluation.There were no further details provided.If significant additional information is received, this report will be supplemented.
 
Event Description
The user found the pin within the connector of the maj-1500 was broken.So it is possible that reprocessing of the endoscopes were insufficient.There was no patient injury report related to the event.
 
Manufacturer Narrative
The subject maj-1500 was not returned to olympus medical systems corp.(omsc) for evaluation.The exact cause of the reported event could not be conclusively determined.Based on device investigation results of similar events reported in the past, the pin was likely broken due to degradation and excessive force.The instruction manual provides of maj-1500 provides warnings and how to inspect the device prior to use as follows.Warning over time, the connecting tube will deteriorate because it is subject to wear with each use.If any of the following are found with the connecting tube, do not use it and replace it with a new one.Connecting a defective tube could prevent the effectiveness of the cleaning and high-level disinfection process.Inspection visually inspect the pin of the connecting tube to ensure that there are no bends or breaks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECTING TUBE
Type of Device
ENDOSCOPE REPROCESSOR ACCESSORIES
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key9155344
MDR Text Key217692851
Report Number8010047-2019-03494
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170233937
UDI-Public04953170233937
Combination Product (y/n)N
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 10/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberMAJ-1500
Device Lot Number2121547
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/10/2019
Initial Date FDA Received10/04/2019
Supplement Dates Manufacturer Received10/08/2019
Supplement Dates FDA Received10/31/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-