Catalog Number EL5ML
Device Problems Failure to Form Staple (2579); Activation, Positioning or Separation Problem (2906); Ejection Problem (4009)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/12/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # unk.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Attempts are being made to retrieve the device.To date, the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported that during a right thoracotomy, the doctor was trying to ligate peripheral vessels and encountered the following issues with the clip applier.Clips were scissoring, clips were falling out of the device, and clips were not deploying.The doctor continued to use device to complete the procedure.There were no patient consequences reported.
 
Manufacturer Narrative
(b)(4).Batch # p9259v.Investigation summary: the analysis results found that the el5ml device was returned with no damage in the external components.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 6 conforming clips.Upon testing, the jaws opened and closed without any difficulties.In addition, the device locked out as intended.No conclusion could be reached as to what may have caused the reported incident.The reported complaint could not be confirmed.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key9156638
MDR Text Key202725899
Report Number3005075853-2019-22577
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 09/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/24/2022
Device Catalogue NumberEL5ML
Device Lot NumberP4RF8E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2019
Date Manufacturer Received10/04/2019
Patient Sequence Number1
-
-