Catalog Number 0475100000 |
Device Problems
Sparking (2595); Noise, Audible (3273)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/10/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
|
Event Description
|
Per additional information received, it was reported that the device made a popping sound and there was a spark.
|
|
Manufacturer Narrative
|
Alleged failure: cib colten, rep, the unit exploded.Po temp the failure(s) identified in the investigation is consistent with the complaint record.The root cause is the rf board.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
|
Event Description
|
Per additional information received, it was reported that the device made a popping sound and there was a spark.
|
|