Information was received from a manufacturer representative regarding a patient who is implanted with a neurostimulator.It was reported that they did a new paddle lead implant on the day of the report and the check connectivity showed two contacts out (could not recall which two.) upon electrode impedance check, only contacts 6 and 7 showed greater than 40,000 ohms.The health care professional pulled out and repositioned the lead several times along with putting the lead into the 8-15 port of the implantable neurostimulator.When in the 8-15 port, 13 and 14 showed 40,000 ohms.It was confirmed that it was the same contacts as prior to swapping the ports.They ran impedances several times, and it was consistently the same contacts in either port that were over 40,000 ohms.The other contacts had normal, in range impedances.When the physician¿s assistant was inserting the lead into the header block, the lead bent slightly on the white marker band on the lead tail.They were not comfortable using the lead as most of the patient¿s pain is on the left side, so a new lead was used with no issues.There were no patient symptoms or complications reported.
|