MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 37601
Device Problems Electromagnetic Interference (1194); Overheating of Device (1437); Insufficient Information (3190)
Patient Problems Burning Sensation (2146); Shaking/Tremors (2515)
Event Date 10/02/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer regarding a patient who was implanted with a neurostimulator (ins) for essential tremor and movement disorders.It was reported that the day prior the rep turned the patient's device off for an mri and then turned it back on after.Caller said that ever since then the device had not been working.Caller said she would adjust the settings but the patient's shaking has not settled down at all and the patient said that the programmer felt warm on their chest.Caller said there was a warning on the stimulator and said the manual told them to synchronize the programmer to the ins.They synced with the programmer and it showed stimulator on.Increasing/decreasing amplitude did not resolve the issue.No further complications were reported or anticipated with this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a healthcare provider reporting that the event was resolved through a representative going in and adjusting the therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key9163248
MDR Text Key184389446
Report Number3004209178-2019-19132
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00643169529786
UDI-Public00643169529786
Combination Product (y/n)N
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Type of Report Initial,Followup
Report Date 10/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/14/2019
Device Model Number37601
Device Catalogue Number37601
Was Device Available for Evaluation? No
Date Manufacturer Received10/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age69 YR
-
-