• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM ONBOARD BATTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM ONBOARD BATTERY Back to Search Results
Model Number 295025-001
Device Problem Battery Problem (2885)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/18/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses low risk to the patient because the reported issue did not prevent the freedom driver from performing its life sustaining functions.The freedom onboard battery will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the freedom onboard battery was unable to maintain a charge for longer than an hour while the freedom driver was supporting a patient.There was no reported adverse patient impact.
 
Manufacturer Narrative
During the investigation, the customer-reported issue could not be confirmed nor reproduced.The freedom onboard battery was able to power a freedom driver for over 60 minutes and there were no signs of a decreased capacity below syncardia's acceptance criteria.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 4975 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM ONBOARD BATTERY
Type of Device
BATTERY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson, az AZ 85713
MDR Report Key9164225
MDR Text Key161539060
Report Number3003761017-2019-00292
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003084
UDI-Public(01)00858000003084
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 03/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number295025-001
Device Catalogue Number295025-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2020
Initial Date Manufacturer Received 09/18/2019
Initial Date FDA Received10/07/2019
Supplement Dates Manufacturer Received09/18/2019
Supplement Dates FDA Received03/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age41 YR
-
-