Model Number 595000-001 |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/19/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because the reported issue did not prevent the freedom driver from performing its life-sustaining functions.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The customer, a syncardia certified hospital, reported that the freedom driver exhibited a "scratching mechanical noise" while supporting the patient.There was no reported adverse patient impact.The customer also reported that the patient was subsequently switched to a backup freedom driver.
|
|
Manufacturer Narrative
|
Visual inspection of the driver revealed no abnormalities.The driver in "as received" condition passed all sections of functional testing.Additionally, an extended observation run was performed on the driver where no unusual noises were produced.The driver performed as intended.During investigation testing, the customer-reported "scratching mechanical noise" was not reproduced and there was no evidence of a device malfunction.The root cause of the customer-reported issue could not be conclusively determined.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
|
|