Catalog Number A1114 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was not yet returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.
|
|
Event Description
|
A customer reported that the a1114 mayfield infinity skull clamp did not clamp all the way down on one side on an unknown date.Additional information received on 17sep2019 indicating that the product problem occurred during application of the mayfield in a craniotomy procedure.There was no harm to the patient.A surgery delay of less than 10-minute was noted.
|
|
Manufacturer Narrative
|
The device was returned for evaluation.The right side boss threaded hole was cross threaded.The unit needed to be machined to have heli-coils added to large starburst threads.The reported complaint was confirmed from the evaluation of item.The definite root cause of the reported complaint could not be reliably determined.
|
|