The subject device was returned for evaluation for the reported "ceramic tip broke off".The evaluation confirmed the ceramic tip was broken off as a visual inspection of the device found the black ceramic insulation tip (beak) was missing and not returned.The remaining portions of the beak inside the distal end has jagged edges indicating the tip had broken off.However, there is evidence of adhesive remains inside the distal sheath which indicates the beak being glued properly.Furthermore, multiple deep dents and scratches were found along the outer tube of the sheath.Mechanically, the device passed inspection as the lock ring, release button function, and passage function appropriately.Based on the evaluation findings, the sharp jagged edges on the remaining beak inside the sheath, including the dents and scratches on the outer tube, are an indication of physical damage due to impact or stress.To mitigate the risk of device tip damage the instruction manual cautions, "if applying excessive force on the instrument during insertion or removal or hitting the tip against other instruments will cause the insulation tip to break." in addition, "if the inner sheath is inserted or removed at an angle to the outer sheath, the lateral force applied to the inner sheath may crack, loosen, break, or otherwise damage the sheath¿s insulated distal tip." as a preventive measure, the instruction manual provides indicates prior to operation, to feel the entire surface of the instrument for dents, protrusions, or other irregularities¿do not use if damage is found.
|