H.10: further analysis on the serial read out revealed that the cardioplegia pump was linked to the master pump by the stop-link option.This means that every time the master pump is stopped by any alarm, the cardioplegia pump stops too.The cardioplegia pump got a stop due to the monitoring of the master pump (arterial pump).Right after this event, a general override was activated on the cardioplegia pump.When the general override is activated the external monitor is disabled for an unlimited period, thus all the alarm functions of the assigned devices are likewise stopped as the pump is stopped.The unit was put back in service.Basing on the above, it is possible to conclude that no hardware / software failure occurred and both pumps worked as expected.The root cause of the reported issue has been addressed back to a user error.
|