MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problems Mechanical Problem (1384); Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 10/01/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses low risk to a patient because the issue was observed when the freedom driver was not supporting a patient.The freedom driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The freedom driver was not supporting a patient.The customer, a syncardia certified hospital, reported that both of the freedom onboard batteries were unable to latch in the freedom driver.
 
Manufacturer Narrative
Visual inspection of the driver revealed fractured plastics, battery safety latch stuck in the pressed down position and a battery safety latch spring out of the correct position.During investigation testing, the customer-reported issue was reproduced as the onboard batteries were unable to latch in the freedom driver.The root cause was determined to be that the right battery safety latch and spring were stuck due to the safety latch spring out of the correct position.It could not be conclusively determined how the spring on the right battery safety latch moved out of the correct position.Once the spring was repositioned correctly, the freedom driver passed all sections of functional testing.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key9176928
MDR Text Key161987165
Report Number3003761017-2019-00296
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 11/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2019
Initial Date Manufacturer Received 10/01/2019
Initial Date FDA Received10/10/2019
Supplement Dates Manufacturer Received10/01/2019
Supplement Dates FDA Received11/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age53 YR
-
-