Model Number 23-07-50
Device Problem False Alarm (1013)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/18/2018
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s3 bubble detector.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and the cushions of the bubble sensor were found to be deflated.The sensor has been replaced and the issue solved.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.
 
Event Description
Livanova (b)(4) received a report that a pump stop has been triggered by a bubble alarm.The perfusionist was not able to start the pump again and resume the bypass.Therefore, the s3 bubble detector was switched off and the perfusionist was able to start the pump again.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S3 BUBBLE DETECTOR
Type of Device
MONITOR AND/OR CONTROL, LEVEL SENSING, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key9178245
MDR Text Key194352332
Report Number9611109-2019-00809
Device Sequence Number1
Product Code KRL
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K955152
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 10/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number23-07-50
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/18/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/21/2015
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-