• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA SJM TRIFECTA VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA SJM TRIFECTA VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number TF-23A
Device Problem Material Split, Cut or Torn (4008)
Patient Problems Chest Pain (1776); Dyspnea (1816); Fatigue (1849); Mitral Regurgitation (1964)
Event Type  Injury  
Manufacturer Narrative
An event of transvalvular regurgitation, shortness of breath, fatigue, and episodes of chest pain was reported.The results of the investigation are inconclusive since the device remains implanted was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
The 23mm trifecta valve was implanted on (b)(6) 2017.On (b)(6) 2019 severe transvalvular aortic regurgitation was noted.The patient noted that over the previous 3 months they had been experiencing shortness of breath, fatigue, and episodes of chest pain.Scans revealed a non-calcific leaflet tear in the trifecta valve.The patient reported that they had ceased taking statin and exchanged their aspirin for asasantin against the physicians advice.The patient's medication was adjusted.It was reported that the site planned to explant and replace the valve.Additional information has been requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM TRIFECTA VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
pamela yip
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key9178824
MDR Text Key161939575
Report Number3008452825-2019-00476
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P100029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/09/2019
Device Model NumberTF-23A
Device Catalogue NumberTF-23A
Device Lot NumberBR00012675
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/16/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/10/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-