Model Number WCD 4000
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/30/2019
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (wires exposed) has been confirmed.Upon investigation the cable connecting the distribution node (dn) to the rear therapy electrodes was pulled from the strain relief, damaging wires in the cable.The root cause for the strained cable was excessive force.No adverse event resulted from the defective electrode belt.
 
Event Description
A us distributor reported that a patient's electrode belt had exposed wires.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
brooke arnold
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key9181343
MDR Text Key167192613
Report Number3008642652-2019-08070
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005012
UDI-Public00855778005012
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 10/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2019
Date Manufacturer Received09/13/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-