The device has been received at coloplast; however the evaluation is not yet complete.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Once our evaluation is complete, a follow-up report will be submitted.
|
This follow-up mdr is created to document the additional event information and evaluation of the returned device.A titan touch pump was received for evaluation.Examination and testing of the returned component revealed a separation within abrasion on the shorter exhaust tube of the pump.Testing revealed this to be a site of leakage.Abrasion was noted on the longer exhaust tube and inlet tube of the pump.Based on recreation of the position of the tubes according to the abrasion pattern, this demonstrates that all pump tubes were overlapping one another while in-vivo.This positioning, in combination with device usage over time, could contribute to sufficient stress(s) to cause a separation through the shorter exhaust tubing at this site.A separation of this type would then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, non-conformances and capas revealed no trends for this lot.
|