DEPUY SYNTHES PRODUCTS LLC BATTERY RECIPROCATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 530.715 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Udi: (b)(4).The actual device has been returned and is currently pending evaluation.Once the investigation has been completed, and if additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported that during an unspecified surgical procedure, it was discovered that the battery reciprocator device was leaking water.It was not reported if there were any delays in the surgical procedure or if a spare device was available for use.There was patient involvement reported.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device evaluation: this device was returned for evaluation.Quality engineering evaluated the device.A visual and functional assessment was performed which determined that there was an unknown liquid coming from the saw head ratchet; failed saw head ratchet check.It was determined that the device would not rotate.Therefore, the reported condition was confirmed.The assignable root cause was determined to be traced to maintenance, which is improper maintenance.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|