Catalog Number 530.715
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).The actual device has been returned and is currently pending evaluation.Once the investigation has been completed, and if additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported that during an unspecified surgical procedure, it was discovered that the battery reciprocator device was leaking water.It was not reported if there were any delays in the surgical procedure or if a spare device was available for use.There was patient involvement reported.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device evaluation: this device was returned for evaluation.Quality engineering evaluated the device.A visual and functional assessment was performed which determined that there was an unknown liquid coming from the saw head ratchet; failed saw head ratchet check.It was determined that the device would not rotate.Therefore, the reported condition was confirmed.The assignable root cause was determined to be traced to maintenance, which is improper maintenance.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY RECIPROCATOR II FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key9181985
MDR Text Key195903044
Report Number8030965-2019-68981
Device Sequence Number1
Product Code GEY
UDI-Device Identifier07611819491854
UDI-Public07611819491854
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 09/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.715
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2019
Date Manufacturer Received11/01/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-